+8615966019173
Surface Thermometer Calibration Device
Home - Products - Temperature Measuring Equipment - Surface Thermometer Calibration Device - DY-BM02 Surface Thermometer Calibration Furnace
With the development of science and technology, contact surface thermometers are widely used in aerospace, energy, electromechanical, chemical, textile, electric power and other fields, so the calibration and measurement needs of surface thermometers are increasing.
With the development of science and technology, contact surface thermometers are widely used in aerospace, energy, electromechanical, chemical, textile, electric power and other fields, so the calibration and measurement needs of surface thermometers are increasing.
DY - BM02 surface thermometer calibration furnace is developed by taian german-american mechanical and electrical equipment co., LTD. Production of the new temperature calibration equipment, this equipment according to the actual demand, expanding to a temperature of 600 ℃, solved the high temperature surface temperature sensor cannot be traced back problem, it is mainly used for calibration of various type and size of the sensor surface, such as thermistors, thin film sensors, surface resistance thermometer, strip temperature sensor, surface thermocouple.
The calibrator rises and drops quickly, from room temperature to 600 ℃ within 40 minutes; The actual test technical indicators meet the standard requirements of "JJF1409-2013 Surface Thermometer Calibration Specifications", and some technical parameters are better than the imported similar products.
● the temperature control system and temperature source furnace system layout;
● for 7 inch touch screen man-machine dialogue interface, human-machine interactive dialog more friendly, more clear, easy to operate;
● with fine alloy new material as heat conducting plate, with high thermal conductivity, good thermal performance, good oxidation resistance, corrosion resistance is strong, good uniformity, etc;
● using uniform heating plate, improve the uniformity and stability of the heat source, to ensure the work area uniformity of temperature field.
Temperature range | ring temperature +5℃~600.0℃ |
Accuracy | ±0.5℃@200 ℃; Plus or minus 1.0 ℃ @ 400 ℃; Plus or minus 1.5 ℃ @ 600 ℃ |
Stability | ±0.2℃@300℃; Plus or minus 0.3 ℃ @ 400 ℃; Plus or minus 1.0 ℃ @ 600 ℃ |
Resolution | 0.1℃ |
Uniformity | ≤0.5℃@100℃; 0.6 ℃ @ 200 ℃ or less; 0.8 ℃ @ 300 ℃ or less; 1.0 ℃ @ 400 ℃ or less; 2.0 ℃ @ 600 ℃ or less |
Heating time | 40min (room temperature to 600℃) |
Cooling time | 60min (600℃ to 100℃) |
Stabilization time | ≤8min |
Measuring surface | ¢100mm |
Power consumption | 1000W |